Abstract
For the determination of salicylaldoxime in environmental water samples, a stable and rapid method with low detection was proposed and established, based on the liquid-liquid extraction-high performance liquid chromatography with ultraviolet detector. Parameters including extraction solvent, ionic strength, solution pH, and extraction pattern were discussed for the optimal quantification of salicylaldoxime-spiked water. When the described method was applied to four spiked water samples, the obtained average extraction recovery rate was found between 87%–107% and relative standard deviation was below 6%. At the same time, good linear relationships were observed for spiked water samples from 0.01 to 10 μg/mL (R2=0.9993). In addition, the detection limit of salicylaldoxime was revealed between 0.003–0.008 μg/mL, which is two orders of magnitude lower than previously reported results. Thus, the presented method may be advantageous for the high-efficiency determination of salicylaldoxime in water samples.
| Original language | English |
|---|---|
| Pages (from-to) | 701-708 |
| Number of pages | 8 |
| Journal | Journal of Central South University |
| Volume | 25 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - 1 Apr 2018 |
| Externally published | Yes |
Keywords
- high-performance liquid chromatography
- liquid-liquid extraction
- salicylaldoxime
Fingerprint
Dive into the research topics of 'Improved determination of salicylaldoxime in water samples by liquid-liquid extraction followed by high performance liquid chromatographic analysis'. Together they form a unique fingerprint.Cite this